Description

Validate the precision and accuracy of combined blood gas and electrolyte analyzers. Eurotrol GAS-ISE provides three physiological relevant levels simulating clinically significant ranges of acid/base and electrolyte balance, respiratory function, glucose, lactate and urea concentration.

Intended Use

Eurotrol GAS-ISE is an assayed quality control material intended for professional use in the verification of the precision and accuracy of combined blood gas and electrolyte analyzers. Eurotrol GAS-ISE is also developed to verify calibration, operating temperature and other performance characteristics.

Analytes

ARTICLE CODE
013.001.010
013.002.010
013.003.010
Level 1
Level 2
Level 3
pH
7.150 - 7.250
7.350 - 7.450
7.550 - 7.650
1
pO2
67.0 - 77.0
96.0 - 110.0
129.0 - 147.0
mmHg
pCO2
53.0 - 63.0
35.0 - 45.0
16.0 - 24.0
mmHg
Na+
117.0 - 123.0
137.0 - 143.0
157.0 - 163.0
mmol/L
K+
2.70 - 3.30
4.20 - 4.80
6.70 - 7.30
mmol/L
Li+
0.45 - 0.55
0.75 - 0.85
1.15 - 1.25
mmol/L
Ca2+
1.75 - 1.95
1.20 - 1.40
0.65 - 0.85
mmol/L
Cl-
77.00 - 83.00
97.00 - 103.00
117.00 - 123.00
mmol/L
tMg
0.4 - 0.6
0.8 - 1.0
1.4 - 1.6
mmol/L
Glu
2.9 - 3.1
5.9 - 6.1
12.4 - 12.6
mmol/L
Lac
7.20 - 7.80
2.70 - 3.30
0.90 - 1.50
mmol/L
Urea
0.50 - 1.10
3.00 - 3.70
7.20 - 7.80
mmol/L
tCa
2.47 - 2.57
1.67 - 1.77
0.63 - 0.73
mmol/L
All ranges per analyte are target ranges for manufacturing of the product. See actual value sheets with lot specific values, which may vary in range widths and mean values due to instrument chosen for value assignments.

Details

Ampoule
2.5 ± 0.1 mL
10 per Box
Unopened 2 - 8 °C
Opened 2 - 8 °C
Unopened 365 days
Opened 30 sec

Documents

Tutorials

CueSee® Online